Gasmet Fcx-Series

All sample cell and detector options of the Dx- and Cx-series analysers are available for the FCX family of analysers. Sample cell temperature can be chosen as 50 °C or 180 °C. Different sample cell path lengths are available allowing the FCX – series to be used in a variety of different applications. Analysis results are transmitted through galvanically isolated analog outputs (4-20 mA or 0-10 V). The number of gases that can be analyzed depends on the model. A relay controlled multipoint sampling system is also available.

 

 

FTIR Process Gas Analyzer

 

The Gasmet FCX-Series features a comprehensive range of high performance FTIR Process Gas Analyzers. The wall mounted Gasmet FCX Process Gas Analyzers offer reliable results even when analyzing complex gas mixtures. The Gasmet FCX Process Gas Analyzers incorporates a Fourier Transform Infrared spectrometer, gas cell, and signal processing electronics, a built-in Industrial Computer and aTFT-display into a compact enclosure.

The analysis results of Gasmet FCX Process Gas Analyzer are transmitted via 4-20mA Analog Output signals or digital outputs (ModBus)

The enclosure has IP 65 protection and the analyser is equipped with Vortex cooling system. The sample cell can be equipped with an optional heating and

temperature control up to 180 °C. Sample cell absorption path length is selected according to the application.

 

General parameters

Measuring principle: Fourier Transform Infrared, FTIR

Number of components: max. 8 components

Response time, T90: < 120 s, depending on the gas flow and measurement time

Operating temperature: 20 ± 20 °C

Climate Control: Vortex

Storage temperature: -20 - 60°C, non condensing

Power supply: 100-115 or 230 V / 50-60 Hz

Power consumption: 500 W

Spectrometer

Resolution: 8 cm-1 or 4 cm-1

Scan frequency: 10 or 4 scans / s

IR-Source: SiC, 1550 K

Detector: Peltier cooled MCT

Beamsplitter: ZnSe

Window material: BaF2 or ZnSe

Wavenumber range: 900 - 4200 cm-1

 

Sample Cell

Structure: Multi-pass or single pass cells. Available from 1 cm up to 9.8 m.

Standard materials: Stainless Steel or 100 % gold or Rhodium coated aluminium

Connectors: Swagelok 6 mm or 1/4"

Sample cell gaskets: Teflon® coated Viton®. Kalrez® available upon special request

Temperature: 180 °C, maximum

Window material: BaF2 or ZnSe or AR coated ZnSe

Measuring parameters

Zero point calibration: 24 hours, calibration with nitrogen (4.0 or higher N2 recommended)

Zero point drift: < 2 % of measuring range per zero point calibration interval

Sensitivity drift: none

Linearity deviation: < 2 % of measuring range

Temperature drifts: < 2 % of measuring range per 10 K temperature change

Pressure influence: 1 % change of measuring value for 1 % sample pressure change. Ambient pressure changes measured and compensated

 

Gas Inlet and Outlet Conditions

Gas temperature: non-condensing, the sample gas temperature should be the same as the sample cell temperature

Flow rate: 120 - 600 l per hour

Gas filtration: filtration of particulates (2μ) required

Sample gas pressure: ambient

 

Electronics

A/D Converter: dynamic range 95 dB

Signal Processor: 32-bit floating point DSP 120 MFLOPS

 

Computer

Processor: 2 GHz Intel Celeron

Memory: 256 MB SDRAM

Hard disk: > 40 GB

Operating system: Windows XP system

Application software: CALCMET for Windows

Outputs: isolated 4-20 mA for each component or digital RS-232 Modbus

Relays: Sample gas

Function alarm

Service request

Concentration alarm

Display: Door mounted 12” TFT display

 

Air Conditioning

GASMET Fcx is equipped with Vortex cooling system. Other cooling systems on request.

Cooling capacity: 270 W (air 7 bar, 21 ºC, dewpoint –40 ºC)

Air consumption: 425 Nl/min, max.

Air quality: pressure 7-10 bar temperature + 43 ºC, max. dewpoint –40 ºC

 

Options

Sample pump: for ambient air up to 50 °C only

Automatic zero valve: up to 50 °C sample temperatures

Remote Control: Modem and software for remote diagnostics

Sample cell temp.: 50 - 180 °C

 

Enclosure

Material: Stainless steel

Dimensions (mm): 800 * 600 * 300

Weight: 50 kg

Protection: IP 65

CE - Label: according to EMI guideline 89/336/EC

 

Documentatie

3475_GASMET FCX-Series Technical Data 2006 v1.pdf Download: 3475_GASMET FCX-Series Technical Data 2006 v1.pdf