For continuous measurement of hydrogen and noble gases, the
thermal conductivity of the sample gas mixture is measured and used to
calculate the concentration. Only binary gas mixtures can be measured directly.
If more than two components are present, the concentration of all other
constituents must be measured separately using other measuring methods. The
CALOMAT 6 can carry out internal cross correction after transmission of the
external measuring results. There is no need for an external computer.
The CALOMAT 6 is notable for its high measuring range
dynamics (e.g. 0-1 % and 0-100% H2, can be parameterized) and a low
T90 time.
The CALOMAT 6 is a high-end analyzer available in a
19" design or in a robust field unit for mounting in the open air. With
its additional monitoring unit, the CALOMAT 6 with field unit is suitable to be
installed in Ex zone 1 or 2. In general, its range of uses includes all areas
from the measurement of gas purity measurement through to use in process
monitoring and controls to insure product quality.
The analyzers can easily be added to an integrated
automation system via RS 485, PROFIBUS DP and PROFIBUS PA interfaces, and can
be parameterized and operated using the PDM tool. For fast and easy
maintenance, the analyzers can also communicate with the Siprom GA software
tool that is integrated into the Ethernet via TCP/IP.
Ex-uitvoeringen
Met behulp
van een optionele purge eenheid is de CALOMAT 6, OXYMAT 6 en ULTRAMAT 6 in
veldbehuizing toe[pasbaar in explosie gevaarlijke omgevingen. Voor zowel meting
in brandbare en niet brandbare gassen.